Model HE-960HS
Measurement method Electrode type (4-electrode method)
Cell constant 1.0/cm
Temperature sensor specifications Resistance thermometer bulb 1,000Ω /0°C, 3,850ppm/0°C
Measuring range Conductivity 0 mS/cm to 200 mS/cm (measuring range of conductivity before temperature compensation: 0 mS/cm to 500 mS/cm)
Temperature 0.0°C to 100.0°C
Concentration conversion HNO3 0% to 5.00% Preprogrammed in the instrument
NaOH 0% to 5.00%
H3PO4 0% to 5.00%
Option 1 0% to 100.00% Conversion formula is defined by user (temperature compensation and concentration conversion)
Option 2 0% to 100.00%
Repeatability 0.000 mS/cm to 2.000 mS/cm Within ±0.5% full scale Using equivalent input for conductivity
0.00 mS/cm to 20.00 mS/cm Within ±0.5% full scale
0.0 mS/cm to 200.0 mS/cm Within ±0.5% full scale
0 mS/cm to 500 mS/cm Within ±1.0% full scale
0.0°C to 100.0°C Within ±0.2°C
Linearity 0.00 mS/cm to 20.00 mS/cm Within ±0.5% full scale Using equivalent input for conductivity
0.0 mS/cm to 200.0 mS/cm Within ±1.0% full scale
0 mS/cm to 500 mS/cm Within ±2.0% full scale
0.0°C to 100.0°C Within ±0.5°C
Display and output range Conductivity 0 mS/cm to 1000 mS/cm
Temperature -30°C to 160°C
Concentration 0% to 5 %: NaOH, HNO3, H3PO4
0% to 100%: Concentration of the component in the optional conversion formula set by user
Transmission output Number of output points: 3
4 mA to 20 mA DC: input/output isolated type
Maximum load resistance: 900 Ω
Transmission output range: Freely selectable within the measurement range
(Negative terminals of each transmission output channel are connected inside and thus have the same electric potential.)
Contact output Number of relay : 5
  • ALARM or ERROR contact R1, R2 and R3
    • Contact type: relay contact, SPST(1a)
    • Contact rating: 240VAC,3A (resistance load)
    • Contact function: Upper or lower ON/OFF alarm on each measurement items, conductivity, concentration, temperature, including delay time and hysteresis.
    • Contact action: Closed when status or error is in the event. Opened when status is normal or power is down.
  • ALARM or ERROR contact R4
    • Contact type: relay contact, SPDT(1c)
    • Contact rating: 240VAC,3A (resistance load)
    • Contact function: Upper or lower ON/OFF alarm on each measurement items, conductivity, concentration, temperature, including delay time and hysteresis.
    • Contact action: Closed when status or error is in the event, opened when status is normal or power is down for C-NO output. C-NC output is opposite action of C-NO.
  • R1 and R2, R3 and R4 share a common terminal independently.
    Self diagnosis contact RF
    • Contact type: relay contact, SPST
    • Contact rating: 24VDC,1A (resistance load)
    • Contact function: Any of error or abnormality or under maintenance, can be listed for diagnosis.
    • Contact action: Closed when status is normal. Opened when any erroneous status is detected or power is down.
Communication output The RS-485 input/output
Calibration function Conductivity: Based on the specified compensation coefficient for the cell constant(parameter input)
Temperature: Calibrated by comparing with the reference thermometer
Transmission output hold feature Selectable from the Previous value hold and the Optional value hold.
(However, only the previous value hold is available in the maintenance mode.)
Self-diagnosis function
  • Sensor diagnosis
    Short-circuit and disconnection of the temperature sensor
  • Out of the measurement range
  • Converter error
Temperature compensation
  • Temperature characteristics of NaCl (reference temperature: 25°C)
  • Arbitrary temperature coefficient entry (reference temperature: 25°C, temperature coefficient: ±5%/°C)
  • No temperature compensation is performed.
Temperature-compensated range 0°C to 100°C(a temperature less than 0°C and more than 100°C is calculated as an extension of the range)
Ambient temperature -5°C to 55°C
Relative humidity 20% to 85% (without dew condensation)
Power supply Rated voltage 100 V to 240 V AC, 50/60 Hz, 15 VA (max.)
Structure Indoor-use panel installation type
Panel case: ABS, Terminal: PBT
Panel: IP65 dust and water proof structure
Protective structure Panel: IP65 (IEC60529, JIS C0920)
Rear case: IP20, Terminal:IP00
Class II device (IEC61010-1)
Pollution level 2 (IEC61010-1)
Conforming standards CE Marking EMC Directive (89/336/EEC)
EN61326: 1997/A3:2003
Low Voltage Directive (73/23/EEC)
EN61010-1: 2001
FCC Rule FCC Part15
External dimensions 96(W) mm x 96 (H) mm x 115 (D) mm
Case depth: approx. 105 mm (when panel-mounted)
Mass Approx. 550 g
Compatible sensors Submersible FES-210L series sanitary four-electrode conductivity sensor
Flow-through FES-310L series sanitary four-electrode conductivity sensor

Related Info
Products

Conductivity Meter / Resistivity Meter

To download catalogues, user registration is required.

Products


Sanitary Conductivity sensor
(Flow Through type)
FES-310L Series

{Name} FES-310L Series
Straight structure without differences in levels or protuberances that can completle drain CIP cleaning solutions.
Sanitary Flow Through
High Concentrations CIP
food & beverage

Sanitary Conductivity sensor
(Ferrule Clamp type)
FES-210L Series

{Name} FES-210L Series
Flat structure keeps dirt from adhering.
Sanitary
Flat structure
Ferrule Clamp
High Concentrations CIP
food & beverage

Conductivity meter
(For High concentrations)
HE-480H
Conductivity meter (For high density) HE-480H
Continuous measurement of conductivity, salinity of salt water, and temperature is possible.
General Purpose
Concentration Control
High Concentrations IP65

Conductivity meter
(For Low concentrations)
HE-480C
Conductivity meter (For low density) HE-480C
Compatible with a wide range of applications from management of boiler water to pure water.
General Purpose
Pure Water IP65
Low Concentrations USP

Sanitary Conductivity Meter
(For Low concentrations)
HE-960CW
2-Channel Conductivity Meter HE-960CW
Ideal for the conductivity control of the pasteurized water, and WFI in pharmaceuticals production
General Purpose
USP Pharmaceuticals
Low Concentrations
4 ranges IP65

TMAH Conductivity Meter
HE-960TM
TMAH Conductivity Meter HE-960TM
Perfect for controlling TMAH developer concentrations in semiconductor and liquid crystal processes.
Semiconductor wet process
Concentration Control
chemical IP65

Conductivity meter
(For high concentrations)
CEH-200
Conductivity meter (For high density) CEH-200
This design fulfills basic specifications while raising the bar in terms of ease of operation.
General Purpose
Concentration Control
High Density

Conductivity meter
(For low concentrations)
CE-200C
Conductivity meter (For low density) CE-200C
Rainproof structure that allows field pure water measurements
General Purpose
Pure Water
Low Concentrations
 
TOP
Copyright (C) 2008 HORIBA ADVANCED TECHNO,Co.,Ltd. All rights reserved.

Copyright © 2008 HORIBA ADVANCED TECHNO,Co.,Ltd. All rights reserved. The information shown on this document may be modified without notice. Refer to the original web page for update. The page was copied from:
 
Explore the future HORIBA Advanced Techno