| Model |
HE-960HS |
| Measurement method |
Electrode type (4-electrode method) |
| Cell constant |
1.0/cm |
| Temperature sensor specifications |
Resistance thermometer bulb 1,000Ω /0°C, 3,850ppm/0°C |
| Measuring range |
Conductivity |
0 mS/cm to 200 mS/cm (measuring range of conductivity before temperature compensation: 0 mS/cm to 500 mS/cm) |
| Temperature |
0.0°C to 100.0°C |
| Concentration conversion |
HNO3 |
0% to 5.00% |
Preprogrammed in the instrument |
| NaOH |
0% to 5.00% |
| H3PO4 |
0% to 5.00% |
| Option 1 |
0% to 100.00% |
Conversion formula is defined by user (temperature compensation and concentration conversion) |
| Option 2 |
0% to 100.00% |
| Repeatability |
0.000 mS/cm to 2.000 mS/cm |
Within ±0.5% full scale |
Using equivalent input for conductivity |
| 0.00 mS/cm to 20.00 mS/cm |
Within ±0.5% full scale |
| 0.0 mS/cm to 200.0 mS/cm |
Within ±0.5% full scale |
| 0 mS/cm to 500 mS/cm |
Within ±1.0% full scale |
| 0.0°C to 100.0°C |
Within ±0.2°C |
| Linearity |
0.00 mS/cm to 20.00 mS/cm |
Within ±0.5% full scale |
Using equivalent input for conductivity |
| 0.0 mS/cm to 200.0 mS/cm |
Within ±1.0% full scale |
| 0 mS/cm to 500 mS/cm |
Within ±2.0% full scale |
| 0.0°C to 100.0°C |
Within ±0.5°C |
| Display and output range |
Conductivity |
0 mS/cm to 1000 mS/cm |
| Temperature |
-30°C to 160°C |
| Concentration |
0% to 5 %: NaOH, HNO3, H3PO4
0% to 100%: Concentration of the component in the optional conversion formula set by user
|
| Transmission output |
Number of output points: 3
4 mA to 20 mA DC: input/output isolated type
Maximum load resistance: 900 Ω
Transmission output range: Freely selectable within the measurement range
(Negative terminals of each transmission output channel are connected inside and thus have the same electric potential.) |
| Contact output |
Number of relay : 5
- ALARM or ERROR contact R1, R2 and R3
- Contact type: relay contact, SPST(1a)
- Contact rating: 240VAC,3A (resistance load)
- Contact function: Upper or lower ON/OFF alarm on each measurement items, conductivity, concentration, temperature, including delay time and hysteresis.
- Contact action: Closed when status or error is in the event. Opened when status is normal or power is down.
- ALARM or ERROR contact R4
- Contact type: relay contact, SPDT(1c)
- Contact rating: 240VAC,3A (resistance load)
- Contact function: Upper or lower ON/OFF alarm on each measurement items, conductivity, concentration, temperature, including delay time and hysteresis.
- Contact action: Closed when status or error is in the event, opened when status is normal or power is down for C-NO output. C-NC output is opposite action of C-NO.
- R1 and R2, R3 and R4 share a common terminal independently.
Self diagnosis contact RF
- Contact type: relay contact, SPST
- Contact rating: 24VDC,1A (resistance load)
- Contact function: Any of error or abnormality or under maintenance, can be listed for diagnosis.
- Contact action: Closed when status is normal. Opened when any erroneous status is detected or power is down.
|
| Communication output |
The RS-485 input/output |
| Calibration function |
Conductivity: Based on the specified compensation coefficient for the cell constant(parameter input)
Temperature: Calibrated by comparing with the reference thermometer |
| Transmission output hold feature |
Selectable from the Previous value hold and the Optional value hold.
(However, only the previous value hold is available in the maintenance mode.) |
| Self-diagnosis function |
- Sensor diagnosis
Short-circuit and disconnection of the temperature sensor
- Out of the measurement range
- Converter error
|
| Temperature compensation |
- Temperature characteristics of NaCl (reference temperature: 25°C)
- Arbitrary temperature coefficient entry (reference temperature: 25°C, temperature coefficient: ±5%/°C)
- No temperature compensation is performed.
|
| Temperature-compensated range |
0°C to 100°C(a temperature less than 0°C and more than 100°C is calculated as an extension of the range) |
| Ambient temperature |
-5°C to 55°C |
| Relative humidity |
20% to 85% (without dew condensation) |
| Power supply |
Rated voltage 100 V to 240 V AC, 50/60 Hz, 15 VA (max.) |
| Structure |
Indoor-use panel installation type
Panel case: ABS, Terminal: PBT
Panel: IP65 dust and water proof structure |
| Protective structure |
Panel: IP65 (IEC60529, JIS C0920)
Rear case: IP20, Terminal:IP00
Class II device (IEC61010-1)
Pollution level 2 (IEC61010-1) |
| Conforming standards |
CE Marking |
EMC Directive (89/336/EEC)
EN61326: 1997/A3:2003
Low Voltage Directive (73/23/EEC)
EN61010-1: 2001 |
| FCC Rule |
FCC Part15 |
| External dimensions |
96(W) mm x 96 (H) mm x 115 (D) mm
Case depth: approx. 105 mm (when panel-mounted) |
| Mass |
Approx. 550 g |
| Compatible sensors |
Submersible FES-210L series sanitary four-electrode conductivity sensor
Flow-through FES-310L series sanitary four-electrode conductivity sensor |