| Model |
HT-100 |
| Measurement method |
Continuous UV oxidation and conductivity |
| Measuring range |
0 to 1000 ppb
(Measuring range : 0 to 100/500/1000 ppb) |
| Display resolution |
0 to 100 ppb |
0.1 ppb |
| 0 to 500 ppb |
| 0 to 1000 ppb |
1 ppb |
| Repeatability |
Within ±2% of the full scale |
| Transmission output |
Number of outputs : 1
DC4 to 20mA / DC0 to 1V (option) : Input/Output isolated type
Max. load resistance : 600Ω (min. load resistance : 100kΩ [with option selected])
Transmission output range : Freely selectable within the measurement range
Output contents : TOC concentration |
| Contact output |
Number of outputs : 2
- Alarm contact output (RLY1)
- Contact type : Relay contact, SPST (1c)
- Contact rating : AC125V 3A, DC24V 3A (resistance load)
- Contact function : Upper limit action (ON/OFF control)
- Output contents : Selectable between TOC concentration, Temperature, Conductivity, Hold, and Error (Operation is reversed when Hold or Error is selected)
- Alarm contact output (RLY2)
- Contact type : Relay contact, SPST (1a)
- Contact rating : AC125V 3A, DC24V 3A (resistance load)
- Contact function : Upper limit action (ON/OFF control)
- Output contents : TOC concentration
|
| Sample conditions |
Conductivity : < 10μS/cm or lower (0 to 100 ppb: < 2μS/cm or lower)
Temperature : 5 to 99°C
Pressure : 0.03MPa to 0.50MPa
Flow rate : 20 to 200mL/min |
| Ambient environment |
Temperature : 5 to 40°C
Humidity : 85% or lower |
| Display functions |
LC with backlight
Display content (TOC concentration, conductivity, temperature, alarm) |
| Structure |
Protection : IP43 ;
Paint color 5PB8/1;
Epoxy coating (indoor-use panel installation type) |
| Power supply |
AC100 to 240V±10% 50/60Hz 80VA (max.) |
| Mass |
13kg |
| Conforming standards |