| Model |
HE-960RW |
| Sensor input |
2-channel |
| Cell constant |
Approx. 0.01/cm |
| Temperature compensation element |
Pt 1000Ω (0°C) |
| Measuring range |
Resistivity : 0 to 2.00 / 20.00 / *100.0 MΩ·cm
(0 to 20.0 / 200.0 / *1000 kΩ·cm)
*; Measurable without temperature conversion |
Temperature : 0 to 100°C
Desalination rate : 0 to 100% |
| Repeatability |
within ±0.1% of the full scale (in equivalent input) |
| Linearity |
within ±0.1% of the full scale (in equivalent input) |
| Transmission output |
2 output points,
Max. load resistance: 900Ω
4 to 20 / 0 to 20 mA DC : input / output isolated type
- Transmission output range : Freely selectable within the measurement range. However, repeatability and linearity will remain accurate to the
separately set measuring range.
(Negative terminals of each transmission output channel are connected
inside and thus have the same electric potential.) |
| Contact output |
4 output points; Alarm contact output (R1,R2,R3,R4)
Contact type : relay contact R1, R2, R3 ;SPST
R4 ; SPDT
Contact rating: 240 V AC 3 A and 30 V DC 3 A
(resistance load)
|
| Cmmunication output
|
RS-485 |
| Calibration function |
Specific resistance
Temperature |
Ultra-pure water Specific
resistance
selection |
18.23 (standard),
18.18, 18.24 MΩ·cm
182.3 (standard),
181.8, 182.4 kΩ·cm
|
| Temperature compensation |
Temperature compensation for ultra-pure water and impurities
Reference temperature: 5°C to 95°C |
| Temperature compensation range |
0 to 100°C |
| Ambient environment |
Temperature: -5°C to 45°C, Relative humidity: 20% to 85% (without dew condensation) |
| Power supply |
100 to 240 V AC ±10%, 50/60 Hz, 15 VA (max) |
| Protective Structure |
Panel : IP65, Rear case : IP20, Terminal : IP00
(Indoor-use panel instrallation type) |
| Weight |
Approx. 550 g |
| Compliance |
CE marking FCC Part15 |