| Model : |
IP-960 |
| Measurement principle : |
Alternating current dual electrode method |
| Measuring range : |
Resistivity
0.00-10.00 GΩ.cm
0-1000 MΩ.cm
0.0-100.0 MΩ.cm
(Each range can be switched with a digital switch.) |
| Measurement and display : |
up to 99.99GΩ.cm
Transmission output range described in items 5 and 6. |
| Reproducibility : |
Within ±5% FS |
| Transmission output : |
4-20 mA, isolated input/output type
Max. load resistance: 500 Ω, selectable linear output (selectable with internal DIP switch) |
| Transmission output range resistivity : |
(Linear): 0.00-10.00 GΩ.cm
0-1000 MΩ.cm
0.0-100.0 MΩ.cm
Resistivity (logarithm, 2 digits)
0.10-10.00GΩ cm
10-1000MΩ cm
1.0-100.0MΩ cm |
| Alarm output : |
1c contact output: 2 independent points
Contact capacity: DC 24 V, 1 A max. (resistance load)
Alarm operation: On/Off
Operation: Upper limit or lower limit operation (selectable with internal DIP switch) |
| Upper / lower limit setting range : |
Within measuring range of each range |
| Control width : |
0-20% (0±10%) of full scale for each range
Settings can be made with absolute values |
| Calibration function : |
According to setting of cell constant |
| Additional functions : |
1) Various error messages
- Sensor diagnostic error (temperature sensor disconnection, temperature sensor short-circuit)
- Calibration error (temperature calibration)
- Outside measuring range error (over measuring range, under measuring range)
2) Temperature calibration function (One-point temperature calibration)
3) Moving average cycle setting function (5-100 cycles) |
| Reference temperature : |
25°C |
| Automatic temperature compensation : |
On: According to IPA characteristics
Off: Display of resistivity for temperature at that time
(Selectable with internal DIP switch) |
| Temperature compensation range : |
10-85°C |
| Temperature compensation accuracy : |
±1%/FS |
| Temperature compensation element : |
Pt 1000 Ω |
| Ambient temperature and humidity : |
5-40°C, 85% RH or less |
| Power supply : |
DC 24 V, 10 W |
| Structure: |
Panel -mounted
Indoor installation Case: SUS304 Panel: ABS |
| External dimensions : |
96(W) x 96(H) x 145(D) |
| Weight : |
Approx. 700 g |
| CE marking : |
Compatible with CE marking |
| Safety : |
Insertion of barrier unit between sensor and meter |