| Model : |
CD-480S |
| Measuring range : |
0-20 mg/L
0-200% (saturation: displayed with parameters)
0-40°C (displayed with parameters) |
| Transmission output range : |
0-5, 0-10, 0-15, 0-20 (mg/L) (selectable with internal DIP switch) |
| Display resolution : |
Concentration: 0.01 (mg/L)
Saturation: 1 (%)
Temperature: 0.1 (°C) |
| Transmission output : |
DC 4-20 mA (isolated input/output type)
Max. load resistance: 500 Ω, linearity: ±0.5% FS |
| Adjustment function : |
1c contact output, 2 CH independent (capacity: AC 250 V, 3 A max., resistance load)
Control operation: On/off, upper limit (Hi) or lower limit (Lo) (selectable with internal DIP switch)
Control width: 0.1-1.0 mg/L |
| Calibration functions : |
Zero calibration, atmospheric calibration, saturated solution calibration, temperature calibration |
| Additional functions : |
1) Measured value shift function
2) Various error messages:
1. Calibration error (zero point, gain error, temperature) 2. Outside measuring range error (over full scale, under full scale)
3) Transmission output during calibration: Continuous, hold (previous value, arbitrary value) |
| Temperature compensation range : |
0-40°C |
| Salinity calibration range : |
0-5% |
| Reproducibility : |
Within ±2%/FS (under same temperature conditions) |
| Stability : |
Within ±2%/FS/week (under same temperature conditions) |
| Ambient temperature and humidity : |
-5 to 40°C, 85% RH or less |
| Power supply : |
AC 90-260 V, 50/60 Hz, 10 W |
| Structure : |
Panel-mounted, DIN 48 x 96
Indoor installation Case: SPCC Panel: ABS |
| External dimensions : |
48(W) x 96(H) x 145(D), case depth: 135 mm |
| Weight : |
Approx. 450 g |
| Oxygen sensors : |
5401, 5400, 5405 |
| Probe : |
DP-100 (probe length: 10 m) |